Se rendre au contenu
RDM

HT1-XS Hot Tack Tester

Référence: RDM-HT-1XS

https://www.analis.com/web/image/product.template/146658/image_1920?unique=ad3a0d8

The HT-1XS Hot Tack Tester measures hot tack strength directly after sealing, a critical factor when developing and checking flexible packaging materials. Built for laboratory use with accuracy, repeatability and ease of use at its core, it delivers consistent results that meet the strictest standards, including ASTM F1921.

An intelligent controller, a user-friendly touchscreen and storable test settings make the instrument fast to run and simple to repeat across operators.

What is hot tack and why test it?

Hot tack is the strength of a seal while it is still hot, immediately after sealing and before it has fully cooled. It determines whether a package holds together under the stress of a filling line, which makes it essential for food and pharmaceutical packaging, seal development at film producers, and R&D into optimal seal parameters.

How does it work?

The HT-1XS uses five independent seal bars to build a full temperature profile of sealability in a single test, each bar set from room temperature to 250 °C. It measures force directly after sealing and shows force-time and force-temperature graphs at once, so you see the complete sealing behavior in one run.

Key advantages

  • Five independent seal bars to build a full temperature profile of sealability in one test, each bar set from room temperature to 250 °C
  • Variable dwell time adjustable from 0.1 to 99.9 seconds for each of the five bars
  • Adjustable seal pressure from 0.14 to 4 psi (0.1–2.76 N/cm²) using mechanical weights
  • Direct force measurement after sealing, with force-time and force-temperature graphs shown at once
  • Intuitive operation via touchscreen or foot switch; test data can be viewed, printed and exported to spreadsheets or SPC systems
  • High reproducibility through high-quality components, accurate PID temperature control and UKAS-traceable calibration; measurement accuracy per ISO 9001

Suitable for

  • Packaging of food and pharmaceutical products
  • Seal development and quality control at film producers
  • Research into optimal seal parameters in R&D

Request a demonstration or quote

Download the brochure below for full specifications. Ask Analis’ local Benelux experts for more information, a demonstration, or a personalized quote.

    This combination does not exist.

    Données Techniques

    Category

    Specification

    Application

    Testing of packaging and materials

    Standards

    ASTM F2029

    Compressed air

    External clean, dry compressed air, max. 7 bar

    Storage temperature

    5 °C to 50 °C

    Relative humidity (RH)

    Not more than 80%, non-condensing

    Sealing temperature

    Ambient temperature to 250 °C

    Sealing pressure

    0.1 N/cm² to 2.76 N/cm²

    Seal area