IKA

HABITAT photo cell cct

Reference: IKA-0010007595

https://www.analis.com/web/image/product.template/110410/image_1920?unique=95a34d1

* may be combined with double-walled vessel packages 

The HABITAT photo cell cct control unit package of the HABITAT research bioreactor is used to monitor and control the bioprocess. With the help of the LED panels included in the scope of delivery, the system can be used as a photobioreactor, for example for algae cultivation. The adjustment of the light intensity enables the simulation of the day-night rhythm. Furthermore, the package includes the control unit with all connection options for the gas supply, liquid addition, sensors and temperature control as well as a tablet for simple and clear operation of the bioreactor. In addition to the HABITAT cell package, you have the option of connecting CO2, conductivity and turbidity sensors. 

Control and monitoring 
The bioprocess and all associated test parameters can be easily controlled and monitored via the tablet attached to the control unit and with the help of the intuitive and easy-to-use software. Depending on the type of cultivation, you can choose between the operating modes batch, fed-batch and perfusion/continuous. The new Chaotic Mixing function ensures faster and more effective mixing when required. The 10.4-inch tablet allows direct access to all actuators directly on the main screen. A clear calibration management, a diagram overview and the complete documentation of the process round off the operation. A software version that meets FDA CFR Part 11 requirements is available as an option. 

Gas supply 
Built in mass flow controllers for 4 separate gas lines (for N2, O2, air and CO2) ensure precise, individually adjustable gassing, ideally tailored to the needs of your cells. Flow rates of 0.01-2 l/min can be achieved. 

Liquid supply 
4 integrated Watson Marlow pumps, adjustable in direction and speed, give you the variability to pump in and pump out different liquids (like acid, base, anti-foam agent, feeding solutions). 

Sensors 
HABITAT cell enables the measurement of the following parameters by means of sensors: 
• pH 
• DO (dissolved oxygen) 
• temperature 
• filling level 
• foam 

Temperature control 
The constant and precise temperature control during the cultivation is guaranteed by the heating sleeve, which is adapted to the respective vessel size. (Heating sleeve is included with the HABITAT cell vessel package). 

All features at a glance 
> compact, space-saving design, dimensions: 223 x 402 x 450 mm (WxDxH) 
> large tablet for clear operation (10,4 inch) 
> intuitive, easy-to-use software with a lot of functionalities. 
> 4 integrated fast-load pumps (Watson Marlow) 
> connectivity: USB, PC, RS232, ethernet, external signal input, external pump, single-use unit, thermostat 
> data storage 
> gas supply with 4 built-in mass flow controllers: e.g. for O2, Luft, N2, CO2 
> Status LED display: direct error display by indicator light 

Note: Only a control tower package and the separately available vessel package make up a functional unit. 

 

    This combination does not exist.

    Technical information

    Controller: Cell Growth
    Operating mode: timer, continuous and program operation
    Display: Tablet PC
    Operation: touch screen
    Data memory size [GB]: 32
    Auto restart after power failure: yes
    Ethernet interface: yes
    Single use reactor interface: yes
    Thermostat interface: yes
    Ext. pump interface: yes
    Filter heater: yes
    LED light panel: yes
    Stirring motor: yes
    Calibration guide tepmerature: yes
    Calibration guide pH: yes
    Calibration guide pO2: yes
    pH - temperature compensation: yes
    Mode hose filling: yes
    Stirrer control mode: off, manual, profile, Cascade-pO2 control, chaotic
    Speed range [rpm]: 1 - 2200
    Speed deviation >= 100 rpm [%]: ±2
    Speed deviation <= 100 rpm [rpm]: ±5
    Speed display: TFT
    Reversible direction of rotation : yes
    Timer: yes
    Timer display: TFT
    Time setting range [min]: 1 - 120
    Heating blanket: yes
    Heat output [W]: 250
    Heat control: TFT
    Temperature control: PID
    Working temperature [°C]: 5 - 80
    Connection for ext. temperature sensor: PT1000
    Temperature display: yes
    Temperature measuring range [°C]: 0 - 100
    Temperature measurement resolution [K]: 0.1
    Temperature control accuracy [°C]: ±0.2
    Temperature limit min. [°C]: 0
    Temperature limit max. [°C]: 110
    pH measurement: yes
    pH control mode: off, auto, profile
    pH value display: TFT
    pH measuring range [pH]: 0 - 14
    pH measurement resolution [pH]: 0.01
    Accuracy of pH measurement [pH]: 0.02
    Conncetion for pO2 sensor: yes
    pO2 control mode: off, manual, auto, profile
    Measurement range pO2 min. [%sat]: 0
    Measurement range pO2 max. [%sat]: 200
    Resolution pO2 measurement [%sat]: 0.01
    Accuracy of pO2 measurement [%sat]: 0.3
    Conncetion for turbidity sensor: yes
    Turbidity control mode: Monitoring
    Conncetion for conductivity sensor: yes
    Conductivity control mode: Monitoring
    Conncetion for CO2 sensor: yes
    CO2 control mode: Monitoring
    Gas in connection Air [mm]: 6
    Gas out connection Air [mm]: 6
    Gas in connection O2 [mm]: 6
    Gas out connection O2 [mm]: 6
    Gas in connection N2 [mm]: 6
    Gas out connection N2 [mm]: 6
    Gas in connection CO2 [mm]: 6
    Gas out connection CO2 [mm]: 6
    Aeration mode: continuous & automatic
    Display flowrate: l/min, cc/min
    Input pressure [mbar]: 200 - 3000
    Pressure out min. [mbar]: 200
    Pressure out max. [mbar]: 1000
    Flowrate Air min. [cc/min]: 0
    Flowrate Air max. [cc/min]: 2000
    Flowrate O2 min. [cc/min]: 0
    Flowrate O2 max. [cc/min]: 2000
    Flowrate N2 min. [cc/min]: 0
    Flowrate N2 max. [cc/min]: 2000
    Flowrate CO2 min. [cc/min]: 0
    Flowrate CO2 min. [cc/min]: 2000
    Measuring and control accuracy of the gas control units [% fsr]: 2
    Peristaltic pump: yes
    Numbers of pump heads: 4
    Pump head function: Acid, Base, AntiFoam, Level
    Pump head: Watson Marlow 114DV
    Number of pump tubes: 4
    Number of pump rollers: 4
    Flow rate min. [ml/min]: 1.0
    Flow rate max. [ml/min]: 270
    Pump Speed min. [rpm]: 100
    Pump Speed max. [rpm]: 400
    Pump reversible direction of rotation: yes
    Pump speed display: TFT
    Inner tube diameter min. [mm]: 0.5
    Inner tube diameter max. [mm]: 4.8
    Tube outer diamter min. [mm]: 4.1
    Tube outer diamter max. [mm]: 8.4
    Tube material : Silicone
    Tube hardness [Shore A]: 50 - 60
    Dimensions (W x H x D) [mm]: 224 x 624 x 405
    Weight [kg]: 14
    Permissible ambient temperature [°C]: 5 - 40
    Permissible relative humidity: 80% (up to 31°C), decreasing linearly to max. 50 (@40°C)
    Protection class according to DIN EN 60529: IP 20
    RS 232 interface: yes
    USB interface: yes
    WiFi Interface: yes
    Voltage [V]: 100 - 240
    Frequency [Hz]: 50/60
    Power input [W]: 300
    Power input max. [W]: 1100
    Power input standby [W]: 10
    Fuse: 2x T5.0A 250V
     