Overslaan naar inhoud
MEMMERT

ICO150 156l CO2 Incubator with standard sterilisation program

Referentie: MEM-ICO150

https://www.analis.com/web/image/product.template/4265/image_1920?unique=168b2b9

Safety at all times: high-end functions for the protection of cell, bacteria and tissue cultures

For labs requiring redundant safety monitoring rather than single-point control, the Memmert ICO150 stands apart with its dual-sensor architecture: two independent Pt100 sensors (DIN Class A, 4-wire circuit) mutually monitor temperature — one automatically taking over in case of error — while dual-beam NDIR technology ensures precise, drift-resistant CO₂ regulation with auto-diagnostic fault detection. Combined with the AutoSAFETY thermal protection system, the ICO150 is built for continuous, unattended operation where sample security cannot be compromised.

Key technical highlights:

  • Dual-beam NDIR CO₂ control: digital electronic regulation with auto-diagnostic system, acoustic fault indication, and barometric pressure compensation
  • Twin Pt100 temperature sensors (DIN Class A, 4-wire): mutual monitoring, automatic error takeover, accuracy ±0.1 K at 37°C
  • AutoSAFETY (ASF): independent over-/undertemperature protection following setpoint at preset tolerance; heating cuts off on overtemperature, cooling activates on undertemperature; visual and acoustic alarm
  • Built-in sterilization program: 60 minutes at 180°C, without removing the sensors
  • ControlCOCKPIT TwinDISPLAY: adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT color displays, 9 language settings (DE, EN, ES, FR, PL, CZ, HU, IT, ZH)
  • SetpointWAIT function: process/program time only starts once the set temperature is reached
  • 6-side full-surface heating concept: multi-function heating on four sides plus additional door and back heating to prevent condensation
  • Ethernet LAN / USB connectivity: AtmoCONTROL software for programming, management and data transfer; program retained in memory in case of power failure
  • Integral auto-diagnostics: continuous fault diagnostics for both temperature and CO₂

    This combination does not exist.

    Technische Informatie

    Specification Detail
    Interior volume 156 L
    Interior dimensions (W x H x D) 560 x 700 x 400 mm (D less 35mm for fan)
    Interior material Stainless steel 1.4301 (ASTM 304), corrosion resistant
    Exterior dimensions (W x H x D) 719 x 1070 x 591 mm (D +56mm door handle)
    Working temperature range 5°C above ambient to +50°C
    Setting temperature range +18°C to +50°C
    Setting accuracy (temperature) 0.1°C
    Temperature variation in chamber at 37°C ±0.3 K
    Temperature fluctuation over time at 37°C ±0.1 K
    Temperature sensors 2x Pt100, DIN Class A, 4-wire circuit
    CO₂ adjustment range 0–20%
    CO₂ setting accuracy 0.1%
    CO₂ variation over time ±0.2%
    CO₂ control technology Dual-beam NDIR, digital electronic
    O₂ adjustment range 1–20%
    O₂ setting accuracy 0.1%
    Humidity (standard, Peltier-limited) 93% rh ±2.5% (with water dish full and inserted)
    Humidity setting accuracy 0.5% rh
    Humidity active control (option K7) 40–97% rh and rh-Off
    Sterilization program 60 min at 180°C (sensors remain in place)
    Max. shelves 10 (2 perforated stainless steel shelves standard)
    Max. load (chamber / per shelf) 120 kg / 15 kg
    Interface Ethernet LAN, USB
    Voltage / load 230V or 115V, 50/60Hz, approx. 1500 W
    Ambient operating temperature 10°C to 35°C
    Ambient humidity max. 70% rh, non-condensing
    Max. installation altitude 2,000 m above sea level
    Overvoltage category II
    Pollution degree 2
    Net / gross weight (with packaging) 90 kg / 116 kg approx.
    Packaging dimensions 800 x 1250 x 800 mm
    Country of origin Federal Republic of Germany
    Customs tariff number 8419 8998
    WEEE-Reg.-No. DE 66812464
    Safety approval Standard units are safety-approved and bear the test marks

    Standard equipment: works calibration certificate (chamber center, +37°C, 5% CO₂); fully insulated stainless steel door with 2-point compression lock; inner glass door with 8mm Ø gas sampling port; CO₂ connection set (hose, coupling, clamp); membrane filter for incoming gas; 1 stainless steel water dish.