- Analytische chemie
- Beschermingskasten
- Biologische beeldvorming
- Bloedplaatjesaggregatie
- Capillaire elektroforese
- Celculturen en fermentatie
- Cellulaire analyse
- Centrifugatie
- Chromatografie (flash & prep)
- DNA / RNA (nucleïnezuren)
- Deeltjeskarakterisatie
- Desinfectie
- Elektrochemie
- Flowcytometrie
- Flowcytometrie celsorting
- Geautomatiseerde Liquid Handling
- Hematologie
- Immuno-hematologie
- Immunodosage
- In-vitrofertilisatie (IVF)
- Kleur-Appearance en fysische testen
- Klinische IT-oplossingen
- Klinische automatisering (TLA)
- Klinische chemie
- Klinische kwaliteitscontroles
- Laboratoriummeubilair & ontwerpstudies
- Massaspectrometers
- Materiaaltesten en -karakterisering
- Metrologie
- Microbiologie (Klinische)
- Microscopie
- Monstersvoorbereiding
- Nefelometrie
- Omgevingscontrole
- Opslag
- POCT
- Pipetteren van vloeistoffen (handmatig)
- Pompen
- Reinigen
- Reometers
- Shakers and stirrers
- Sterilisatie
- Synthese
- Tablet testing
- Texturometers
- Toxicologie
- Urineanalyse
- Verwarmen & koelen
- Viscometers
- Vochtanalyse
- Vriesdrogen
- Waterzuivering
- Wegen
- Verwarmen & koelen Incubatie CO2-incubatoren
- IC050 56l CO2 incubator with standard sterilisation program
HOME / APPLICATIONS & TECHNIQUES /
IC050 56l CO2 incubator with standard sterilisation program
Referentie: MEM-ICO50
https://www.analis.com/shop/mem-ico50-ic050-56l-co2-incubator-with-standard-sterilisation-program-4061 https://www.analis.com/web/image/product.template/4061/image_1920?unique=4e630bcThe smallest ICO, without compromise on safety
Where bench space is limited but reliability cannot be, the Memmert ICO50 brings the full safety architecture of the ICO range — dual-sensor temperature monitoring, dual-beam NDIR CO₂ control, and AutoSAFETY protection — into a compact 56 L footprint. It is the entry model of a four-size range (56 to 241 L), designed for labs running smaller batches of cell, bacteria or tissue cultures without sacrificing the redundant monitoring found on larger ICO models.
Key technical highlights:
- Dual-beam NDIR CO₂ control: digital electronic regulation with auto-diagnostic system, acoustic fault indication, and barometric pressure compensation
- Twin Pt100 temperature sensors (DIN Class A, 4-wire): mutual monitoring, automatic error takeover, accuracy ±0.1 K at 37°C
- AutoSAFETY (ASF): independent over-/undertemperature protection following setpoint at preset tolerance; automatic heating/cooling cutoff; visual and acoustic alarm
- Built-in sterilization program: 60 minutes at 180°C, without removing the sensors
- ControlCOCKPIT TwinDISPLAY: adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT color displays, 9 languages (DE, EN, ES, FR, PL, CZ, HU, IT, ZH)
- SetpointWAIT function: process/program time only starts once the set temperature is reached
- 6-side full-surface heating concept: prevents condensation even in the smallest chamber of the range
- Ethernet LAN / USB connectivity: AtmoCONTROL software for programming, management and data transfer; program retained in memory in case of power failure
- Mobile alarm notifications: visual and acoustic alarm plus remote alert if temperature, CO₂, O₂ or humidity ranges are exceeded
- 3 years worldwide guarantee
Technische Informatie
| Specification | Detail |
| Interior volume | 56 L |
| Interior dimensions (W x H x D) | 400 x 425 x 330 mm (D less 35mm for fan) |
| Interior material | Stainless steel 1.4301 (ASTM 304), corrosion resistant |
| Exterior dimensions (W x H x D) | 559 x 795 x 521 mm (D +56mm door handle) |
| Working temperature range | 5°C above ambient to +50°C |
| Setting temperature range | +18°C to +50°C |
| Setting accuracy (temperature) | 0.1°C |
| Temperature variation in chamber at 37°C | ±0.3 K |
| Temperature fluctuation over time at 37°C | ±0.1 K |
| Temperature sensors | 2x Pt100, DIN Class A, 4-wire circuit |
| CO₂ adjustment range | 0–20% |
| CO₂ setting accuracy | 0.1% |
| CO₂ variation over time | ±0.2% |
| CO₂ control technology | Dual-beam NDIR, digital electronic |
| O₂ adjustment range | 1–20% |
| O₂ setting accuracy | 0.1% |
| Humidity (standard, Peltier-limited) | 93% rh ±2.5% (with water dish full and inserted) |
| Humidity setting accuracy | 0.5% rh |
| Humidity active control (option K7) | 40–97% rh and rh-Off |
| Sterilization program | 60 min at 180°C (sensors remain in place) |
| Max. shelves | 5 (1 perforated stainless steel shelf standard) |
| Max. load (chamber / per shelf) | 75 kg / 15 kg |
| Interface | Ethernet LAN, USB |
| Voltage / load | 230V or 115V, 50/60Hz, approx. 1100 W |
| Ambient operating temperature | 10°C to 35°C |
| Ambient humidity | max. 70% rh, non-condensing |
| Max. installation altitude | 2,000 m above sea level |
| Overvoltage category | II |
| Pollution degree | 2 |
| Net / gross weight (with packaging) | 55 kg / 74 kg approx. |
| Packaging dimensions | 730 x 950 x 640 mm |
| Country of origin | Federal Republic of Germany |
| Customs tariff number | 8419 8998 |
| WEEE-Reg.-No. | DE 66812464 |
| Safety approval | Standard units are safety-approved and bear the test marks |
Standard equipment: works calibration certificate (chamber center, +37°C, 5% CO2); fully insulated stainless steel door with 2-point compression lock; inner glass door with 8mm Ø gas sampling port; CO2 connection set (hose, coupling, clamp); membrane filter for incoming gas; 1 perforated stainless steel shelf; 1 stainless steel water dish.
Beschermingskasten
Biologische beeldvorming
Bloedplaatjesaggregatie
Capillaire elektroforese
Celculturen en fermentatie
Cellulaire analyse
Centrifugatie
Chromatografie (flash & prep)
DNA / RNA (nucleïnezuren)
Deeltjeskarakterisatie
Desinfectie
Elektrochemie
Flowcytometrie
Flowcytometrie celsorting
Geautomatiseerde Liquid Handling
Hematologie
Immuno-hematologie
Immunodosage
In-vitrofertilisatie (IVF)
Kleur-Appearance en fysische testen
Klinische IT-oplossingen
Klinische automatisering (TLA)
Klinische chemie
Klinische kwaliteitscontroles
Laboratoriummeubilair & ontwerpstudies
Massaspectrometers
Materiaaltesten en -karakterisering
Metrologie
Microbiologie (Klinische)
Microscopie
Monstersvoorbereiding
Nefelometrie
Omgevingscontrole
Opslag
POCT
Pipetteren van vloeistoffen (handmatig)
Pompen
Reinigen
Reometers
Shakers and stirrers
Sterilisatie
Synthese
Tablet testing
Texturometers
Toxicologie
Urineanalyse
Verwarmen & koelen
Viscometers
Vochtanalyse
Vriesdrogen
Waterzuivering
Wegen