Skip to Content
BINDER

CB260-230V-O, with O2 control 0.2–20%, Series CB - CO2 incubators with hot air sterilization and heat sterilizable CO2 sensor, 267 L, 230 V 1~ 50/60 Hz

Reference: BIN-9640-0014

https://www.analis.com/web/image/product.template/127945/image_1920?unique=fb021fe

Scale up your research without scaling down safety

When cell-based assays require larger batches or parallel workflows, the BINDER CB 260 delivers the same rigorous contamination control as the CB range, in a 267 L chamber built for higher-volume protocols. Its heat sterilizable CO₂ sensor and proven 180°C hot air sterilization concept ensure that increased capacity never comes at the cost of reproducibility.

Key technical highlights:

  • 267 L high-capacity chamber: seamless, fanless stainless steel interior designed for multi-sample throughput
  • 180°C hot air sterilization: eliminates contaminants without chemical consumables, validated anti-contamination concept
  • Heat sterilizable CO₂ sensor: single-beam infrared technology, stable measurement with no drift over repeated cycles
  • Double-pan humidification with condensation protection, maintaining 90–95% RH across the full chamber volume
  • Fail-safe CO₂ regulation: protects culture pH stability even during high-frequency door openings
  • Cell therapy compartmentalization option: divided inner door for segregated multi-user or multi-project handling
  • Full traceability: internal data logging with USB export, Ethernet connectivity, zero-voltage alarm contact

    This combination does not exist.

    Technical information

    Specification Detail
    Interior volume 267 L
    Temperature range +5°C above ambient to 50°C
    Temperature uniformity at 37°C ±0.4 K
    Temperature fluctuation at 37°C ±0.1 K
    Recovery time after door opening (30 sec, 37°C) 6 min
    Humidity 90–95% r.h.
    CO₂ range 0–20 vol.%
    CO₂ measurement Infrared (IR), single-beam
    CO₂ recovery time (30 sec door opening, 5 vol.%) 5 min
    Interior dimensions (W x H x D) 620 x 750 x 575 mm
    Exterior dimensions (W x H x D) 740 x 1,020 x 785 mm
    Net weight 125 kg
    Max. shelf load 10 kg
    Max. total load 40 kg
    Number of shelves (standard/max) 3 / 8
    Nominal power 1.5 kW
    Rated voltage 200–230 V, 50/60 Hz

    Available options: O₂ control, CELLROLL roller bottle system, divided inner door for cell therapy compartmentalization, quick sample access viewing window, stacking frame/adapter, shaker (Professional/Universal), APT-COM™ 4 BASIC-Edition monitoring software.