Skip to Content
BINDER

CB170-230V-D, with Access port 30 mm left, Series CB - CO2 incubators with hot air sterilization and heat sterilizable CO2 sensor, 170 L, 230 V 1~ 50/60 Hz

Reference: BIN-9640-0027

https://www.analis.com/web/image/product.template/127951/image_1920?unique=609a893

Exceptional protection for your cell cultures

The BINDER CB 170 is a powerful and versatile CO₂ incubator suitable for cell-based assays in drug and medical research. It is even suitable for complex cultivation experiments or individual environments with hypoxic conditions thanks to its comprehensive assortment of options and accessories.

Key technical highlights:

  • Proven anti-contamination concept: 180 °C hot air sterilization eliminates contaminants without consumables
  • Heat sterilizable CO₂ sensor: single-beam infrared technology for reliable, drift-free measurement
  • Fail-safe CO₂ system: continuous protection of cell culture pH
  • Optimized humidification: double-pan water system with condensation protection
  • Fanless interior: seamless stainless steel chamber with flanges as shelf support system, Venturi CO₂ gas mixing nozzle
  • Intuitive touchscreen controller with fault diagnosis system (visual and acoustic alarm)
  • Integrated data logging: internal recording, read out in open format via USB interface
  • Interfaces: Ethernet, USB, zero-voltage alarm contact

    This combination does not exist.

    Technical information

    Specification Detail
    Interior volume 170 L
    Internal dimensions (W x H x D) 560 x 600 x 505 mm
    Housing dimensions (W x H x D) 680 x 870 x 715 mm
    Net weight (empty) 96 kg
    Permitted load 30 kg
    Load per rack 10 kg
    Number of shelves (std./max.) 2 / 6
    Temperature range +6 °C above ambient to 50 °C
    Temperature variation at 37°C ± 0.3 K
    Temperature fluctuation at 37°C ± 0.1 K
    Recovery time (30s door opening, 37°C) 6 min
    Humidity range 90–95% RH
    CO₂ range 0–20 vol.%
    CO₂ measurement Infrared (IR)
    Nominal power (230 V) 1.3 kW

    Available options: Access port (30 mm, back/left/right), O₂ control 0.2–20%, door hinged on the left, CELLROLL roller bottle set, shaker "Professional"/"Universal", stacking adapter, calibration certificates, IQ/OQ/PQ documentation, analog output 4-20 mA.